| |
Links in this
section:
Introduction Fundamentals of the TEM technique Beam-sample interaction The Analytical TEM Detector Protection Qualitative Analysis Quantitative Analysis Microanalysis Examples (1) Microanalysis Examples (2) Microanalysis Examples (3) Summary
|
Basics of electron beam-sample interaction in the TEM
While the overall concept remains the same, Energy Dispersive X-ray Analysis
in Transmission Electron Microscopy reveals some marked differences from EDS
analysis in the Scanning Electron Microscope (SEM). Normally, SEM samples
are thick enough for the focussed probe to be contained within the sample,
i.e. complete absorption of the primary beam. This causes the beam to spread
within the material and leads to a number of different phenomena that affect
the analytical treatment of raw X-ray counts obtained from the material.
These include atomic number (Z), absorption (A) and fluorescence (F)
effects, which are dealt with by applying matrix corrections. The incident
beam energy, material density and take off angle therefore have a profound
affect on X-ray intensity and quantitative results in the SEM.
In contrast, TEM samples should be thin enough to be beam transparent, so
the ionization volume formed by a focussed probe is much smaller and,
therefore, there is much less electron scattering (Figure 2). High beam
energies (commonly between 100kV - 400kV) utilized in the analytical TEM
also reduce beam scattering. However, due to the relatively small ionization
volume, X-ray generation is much lower than for thick samples. When
analyzing thicker samples in TEM, absorption corrections may be made if the
material density and thickness at the point of analysis are known, whereas
these corrections are not always necessary in thin samples.
Back | Next
|